RIOT/tests/drivers/ph_oem/main.c

289 lines
7.7 KiB
C

/*
* Copyright (C) 2019 University of Applied Sciences Emden / Leer
*
* This file is subject to the terms and conditions of the GNU Lesser
* General Public License v2.1. See the file LICENSE in the top level
* directory for more details.
*/
/**
* @ingroup tests
* @{
*
* @file
* @brief Test application for the Atlas Scientific pH OEM sensor driver
*
* @author Igor Knippenberg <igor.knippenberg@gmail.com>
*
* @}
*/
#include <stdio.h>
#include "xtimer.h"
#include "event/callback.h"
#include "ph_oem.h"
#include "ph_oem_params.h"
#include "ph_oem_regs.h"
#define SLEEP_SEC (5)
/* calibration test is off by default, so it won't reset your previous calibration */
#define CALIBRATION_TEST_ENABLED (false)
static void reading_available_event_callback(event_t *event);
static ph_oem_t dev;
static event_queue_t event_queue;
static event_t event = { .handler = reading_available_event_callback };
static void reading_available_event_callback(event_t *event)
{
(void)event;
uint16_t data;
puts("\n[EVENT - reading pH value from the device]");
/* stop pH sensor from taking further readings*/
ph_oem_set_device_state(&dev, PH_OEM_STOP_READINGS);
/* reset interrupt pin in case of falling or rising flank */
ph_oem_reset_interrupt_pin(&dev);
ph_oem_read_ph(&dev, &data);
printf("pH value raw: %d\n", data);
ph_oem_read_compensation(&dev, &data);
printf("pH reading was taken at %d Celsius\n", data);
}
static void interrupt_pin_callback(void *arg)
{
puts("\n[IRQ - Reading done. Writing read-event to event queue]");
(void)arg;
/* Posting event to the event queue. Main is blocking with "event_wait"
* and will execute the event callback after posting */
event_post(&event_queue, &event);
/* initiate new reading with "ph_oem_start_new_reading()" for this callback
to be called again */
}
int main(void)
{
uint16_t data;
puts("Atlas Scientific pH OEM sensor driver test application\n");
printf("Initializing pH OEM sensor at I2C_%i, address 0x%02x...",
PH_OEM_PARAM_I2C, PH_OEM_PARAM_ADDR);
if (ph_oem_init(&dev, ph_oem_params) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Turning LED off... ");
if (ph_oem_set_led_state(&dev, PH_OEM_LED_OFF) == PH_OEM_OK) {
puts("[OK]");
/* Sleep 2 seconds to actually see it turning off */
xtimer_sleep(2);
}
else {
puts("[Failed]");
return -1;
}
printf("Turning LED on... ");
if (ph_oem_set_led_state(&dev, PH_OEM_LED_ON) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
/* Test changing the pH OEM i2c address to 0x66 and back to 0x65 in the
* sensor as well as dev->params.addr
*/
printf("Setting device address to 0x66... ");
if (ph_oem_set_i2c_address(&dev, 0x66) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Setting device address back to the default address 0x65... ");
if (ph_oem_set_i2c_address(&dev, 0x65) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
/* Test calibration process and if it is applied correctly in the pH OEM register */
if (CALIBRATION_TEST_ENABLED) {
printf("Clearing all previous calibrations... ");
if (ph_oem_clear_calibration(&dev) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Reading calibration state, should be 0... ");
if (ph_oem_read_calibration_state(&dev, &data) == PH_OEM_OK
&& data == 0) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
/* Don't forget to provide temperature compensation for the calibration */
printf("Setting temperature compensation to 22 Celsius... ");
if (ph_oem_set_compensation(&dev, 2200)) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
/* Always start with mid point when doing a new calibration */
printf("Calibrating to midpoint... ");
if (ph_oem_set_calibration(&dev, 6870, PH_OEM_CALIBRATE_MID_POINT)
== PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Reading calibration state, should be 2... ");
if (ph_oem_read_calibration_state(&dev, &data) == PH_OEM_OK
&& data == 2) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Calibrating to lowpoint... ");
if (ph_oem_set_calibration(&dev, 4000, PH_OEM_CALIBRATE_LOW_POINT)
== PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Reading calibration state, should be 3... ");
if (ph_oem_read_calibration_state(&dev, &data) == PH_OEM_OK
&& data == 3) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Calibrating to highpoint... ");
if (ph_oem_set_calibration(&dev, 9210, PH_OEM_CALIBRATE_HIGH_POINT)
== PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
printf("Reading calibration state, should be 7... ");
if (ph_oem_read_calibration_state(&dev, &data) == PH_OEM_OK
&& data == 7) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
}
if (gpio_is_valid(dev.params.interrupt_pin)) {
/* Setting up and enabling the interrupt pin of the pH OEM */
printf("Enabling interrupt pin... ");
if (ph_oem_enable_interrupt(&dev, interrupt_pin_callback,
&data) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
/* initiate an event-queue. An event will be posted by the
* "interrupt_pin_callback" after an IRQ occurs. */
event_queue_init(&event_queue);
}
else {
puts("Interrupt pin undefined");
}
printf("Setting temperature compensation to 22 °C... ");
if (ph_oem_set_compensation(&dev, 2200) == PH_OEM_OK) {
puts("[OK]");
}
else {
puts("[Failed]");
return -1;
}
while (1) {
puts("\n[MAIN - Initiate reading]");
/* blocking for ~420ms till reading is done if no interrupt pin defined */
ph_oem_start_new_reading(&dev);
if (gpio_is_valid(dev.params.interrupt_pin)) {
/* when interrupt is defined, wait for the IRQ to fire and
* the event to be posted, so the "reading_available_event_callback"
* can be executed after */
event_t *ev = event_wait(&event_queue);
ev->handler(ev);
}
if (!gpio_is_valid(dev.params.interrupt_pin)) {
if (ph_oem_read_ph(&dev, &data) == PH_OEM_OK) {
printf("pH value raw: %d\n", data);
}
else {
puts("[Reading pH failed]");
}
if (ph_oem_read_compensation(&dev, &data) == PH_OEM_OK) {
printf("pH reading was taken at %d Celsius\n", data);
}
else {
puts("[Reading compensation failed]");
}
}
xtimer_sleep(SLEEP_SEC);
}
return 0;
}